Two simple methods of using sophisticated statistical techniques to analyse intralaboratory quality control data are described. One method uses a plastic grid to assess the cumulative sum (cusum) graph. An equivalent procedure, which may be carried out by a laboratory computer, gives automated on-line quality control of the Coulter S blood counter. The possible development of simple apparatus to carry out the same function is discussed.