Desenvolvimento de um dispositivo de baixo custo para medidas por quimiluminescência
AUTOR(ES)
Borges, Eduardo P., Fernandes, Elizabeth N., Rocha, Fábio R. P., Reis, Boaventura F.
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2002-12
RESUMO
A simple and low cost device (ca. US$ 150) that comprises two photodiodes fixed in lab-made Perspex flow cell is proposed for chemiluminescence measurements. The characteristics of the device (large observation window and reduced thickness) allow maximizing the amount of the emitted radiation detected. A sensitivity improvement of ca. 50 % was observed by employing two photodiodes for signal measurements. The performance of the device was assessed by the oxidation of luminol by hydrogen peroxide, yielding a linear response within the range of 2.50 to 500 µmol L-1 H2O2. The detection limit was estimated as 0.8 µmol L-1 hydrogen peroxide which is comparable with those obtained by using equipments based on photomultipliers.
Documentos Relacionados
- Um sistema automático de baixo custo para medidas de intervalos de tempo
- Desenvolvimento de um servidor de arquivo para cliente de baixo custo
- Desenvolvimento de um rastreador de baixo custo para sistemas de energia fotovoltáicos
- Desenvolvimento de substrato de Si de baixo custo para célula fotovoltaica, por aspersão térmica a plasma
- Desenvolvimento de um sistema de análise por injeção em fluxo utilizando materiais alternativos de baixo custo para fins didáticos