Determinação potenciométrica em fluxo de cloreto de cetilpiridinio em desinfectantes bucais
AUTOR(ES)
Baptista, Paula C. S., Araújo, Alberto N., Montenegro, Maria da Conceição B. S. M.
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2003-08
RESUMO
The work describes a new procedure for cetylpyridinium chloride determination in oral disinfectants, based on a flow-injection system with potentiometric detection. The determination was based on the measurement of picrate concentration decrease as result of ion-pair reaction with the analyte present in the injected sample. In the optimised set-up the sample injection volume was kept at 400 µL and merged downstream with the reagent solution containing 1,0 x10-5 mol/L of picrate adjusted to pH 5.0 with citrate/citric acid buffer. The flow rate was fixed at 8 mL/min and the reactor length at 40 cm. The proposed procedure enables the determination of cetylpyridinium in the analytical range of 5,0x10-6 - 7,5x10-5 mol/L at a sampling rate of 60/h. The results for real samples had a precision better than 3% and were comparable to the labelled values.