Determination of Cardiac Output and Ejection Fraction with the Dual Cardiac Probe
AUTOR(ES)
Cardenas, Carlos G.
RESUMO
Cardiac output and left ventricular ejection fraction were determined noninvasively at the bedside in 26 patients by using a dual scintillation probe. The probe is a nonimaging detector that records a high frequency time-activity curve of the passage of an intravenously injected radioactive bolus through the heart. Results were correlated with ejection fraction measured by biplane cineangiography (r = 0.80) and cardiac output determined by green dye dilution (R = 0.86). It is concluded that the dual probe provides an accurate noninvasive means of measuring these parameters, and that it may be particularly applicable to serial measurements in patients in the intensive care unit.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=341600Documentos Relacionados
- Simple and safe bedside method for serial measurement of left ventricular ejection fraction, cardiac output, and pulmonary blood volume.
- Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants.
- Quantitative radionuclide angiocardiography. Determination of left ventricular ejection fraction in children.
- Determination of cardiac output by Doppler echocardiography.
- THE EAR OXIMETER: ITS USE CLINICALLY AND IN THE DETERMINATION OF CARDIAC OUTPUT