Determination of etoricoxib in human plasma using automated on-line solid-phase extraction coupled with LC-APCI/MS/MS
AUTOR(ES)
Dalmora, Sérgio Luiz, Brum Junior, Liberato, Ferretto, Ricardo Machado, Oliveira, Paulo Renato de, Barth, Thiago, Sangoi, Maximiliano da Silva
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2008
RESUMO
A liquid chromatography-tandem mass spectrometry method with atmospheric pressure chemical ionization (LC-APCI/MS/MS) was validated for the determination of etoricoxib in human plasma using antipyrin as internal standard, followed by on-line solid-phase extraction. The method was performed on a Luna C18 column and the mobile phase consisted of acetonitrile:water (95:5, v/v)/ammonium acetate (pH 4.0; 10 mM), run at a flow rate of 0.6 mL/min. The method was linear in the range of 1-5000 ng/mL (r²>0.99). The lower limit of quantitation was 1 ng/mL. The recoveries were within 93.72-96.18%. Moreover, method validation demonstrated acceptable results for the precision, accuracy and stability studies.
Documentos Relacionados
- ON-LINE SOLID PHASE EXTRACTION COUPLED WITH LIQUID CHROMATOGRAPHY TANDEM MASS SPECTROMETRY FOR THE DETERMINATION OF CODEINE IN HUMAN PLASMA
- Liquid chromatographic determination of amoxicillin concentrations in bovine plasma by using a tandem solid-phase extraction method.
- Determination of fipronil in bovine plasma by solid-phase extraction and liquid chromatography with ultraviolet detection
- Liquid chromatographic determination of ampicillin in bovine and dog plasma by using a tandem solid-phase extraction method.
- Solid Phase Extraction to On-Line Preconcentrate Trace Cadmium Using Chemically Modified Nano-Carbon Black with 3-Mercaptopropyltrimethoxysilane