Evaluation of Sensititre Automated Reading and Incubation System for Automated Reading of Sensititre Broth Microdilution Susceptibility Plates
AUTOR(ES)
Chapin, Kimberle C.
FONTE
American Society for Microbiology
RESUMO
The Sensititre Automated Reading and Incubation System (ARIS) performs automated reading of Sensititre identification and susceptibility plates. ARIS susceptibility results were compared to manually read results. A total of 212 isolates (3,352 organism-antimicrobial agent combinations) yielded 99.4% essential agreement and categorical error rates of 0.2% minor, 0% major, and 1.0% very major. ARIS yields accurate susceptibility results compared to plates read manually.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=344475Documentos Relacionados
- Validation of the Automated Reading and Incubation System with Sensititre Plates for Antimicrobial Susceptibility Testing
- Direct Susceptibility Testing of Positive Blood Cultures by Using Sensititre Broth Microdilution Plates
- Effects of incubation temperature, inoculum size, and time of reading on broth microdilution susceptibility test results for amphotericin B fgainst Fusarium.
- Sensititre autoreader for same-day breakpoint broth microdilution susceptibility testing of members of the family Enterobacteriaceae.
- Automated reading of MIC microdilution trays containing fluorogenic enzyme substrates with the Sensititre Autoreader.