Evaluation of Serodia Myco II particle agglutination test for detecting Mycoplasma pneumoniae antibody: comparison with mu-capture ELISA and indirect immunofluorescence.
AUTOR(ES)
Barker, C E
RESUMO
The Serodia Myco II particle agglutination test, which the manufacturers claim exclusively detects IgM antibody, was compared with two IgM-specific tests, a mu-capture ELISA, and indirect immunofluorescence for their ability to detect recent Mycoplasma pneumoniae infection. In general there was good agreement among the three tests, all three having similar sensitivity. One hundred and nine (78%) of serum samples gave concordant results in all three assays. Several sera gave positive particle agglutination titres, however, while being negative by the two other assays, and the Serodia Myco II test may not be as specific for detecting M pneumoniae IgM as the other two tests. While the Serodia Myco II test may be a good screening assay, it is unlikely to be a definitive test for M pneumoniae IgM, but may be better than the complement fixation test, particularly in younger patients in whom M pneumoniae IgM is found more frequently.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=502302Documentos Relacionados
- A novel mu-capture enzyme-linked immunosorbent assay based on recombinant proteins for sensitive and specific diagnosis of hemorrhagic fever with renal syndrome.
- Comparison of the Serodia Treponema pallidum Particle Agglutination, Captia Syphilis-G, and SpiroTek Reagin II Tests with Standard Test Techniques for Diagnosis of Syphilis
- Comparison of the Serodia Treponema pallidum Particle Agglutination, Captia Syphilis-G, and SpiroTek Reagin II Tests with Standard Test Techniques for Diagnosis of Syphilis
- Comparison of the indirect fluorescent antibody test with agglutination, complement-fixation, and Coombs tests for Brucella antibody
- Application of an indirect immunofluorescence test for detection of Mycoplasma pneumoniae in respiratory exudates.