Evaluation of the Autobac 1 Susceptibility Testing System in a Clinical Diagnostic Laboratory
AUTOR(ES)
Mogyoros, Miguel
RESUMO
The Autobac 1 system was compared to the disk diffusion and agar dilution methods. Overall, there was good correlation; however, certain antibiotic/bacterial combinations gave unreliable results. In particular, unreliable results were obtained against staphylococci with penicillins, aerobic gram-negative rods with cephalothin, and enterococci with cephalothin and clindamycin.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=352061Documentos Relacionados
- Laboratory evaluation of amikacin susceptibility testing by the AutoBac I system.
- Evaluation of cefamandole susceptibility testing of Enterobacteriacea by the autobac 1 system.
- Laboratory evaluation of the Anderson Technical semiautomated susceptibility testing system.
- Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study.
- Laboratory Evaluation of a Rapid, Automated Susceptibility Testing System: Report of a Collaborative Study