Evaluation of the sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative study.
AUTOR(ES)
Jones, R N
RESUMO
The Sensititre broth microdilution antibiotic susceptibility testing system was found to be comparable to a reference broth dilution method. Of 3,194 minimum inhibitory concentration data pairs, 93.9% were within the acceptable +/- 1-log2 dilution range. Only rare (0.6%) very major errors were found between methods.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=273418Documentos Relacionados
- Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study.
- Evaluation of the Sceptor microdilution antibiotic susceptibility testing system: a collaborative investigation.
- Clinical laboratory evaluation of the Abbott MS-2 automated antimicrobial susceptibility testing system: report of a collaborative study.
- Rapid and overnight microdilution antibiotic susceptibility testing with the Sensititre Breakpoint Autoreader system.
- Evaluation of Sensititre Automated Reading and Incubation System for Automated Reading of Sensititre Broth Microdilution Susceptibility Plates