Fluorescence immunoassay and passive latex agglutination as alternatives to hemagglutination inhibition for determining rubella immune status.
AUTOR(ES)
Fayram, S L
RESUMO
Three different assays for detection of rubella antibodies, hemagglutination inhibition (HAI), fluorescence immunoassay (FIA), and passive latex agglutination (PLA) were used to test 297 human serum samples. Overall agreements for immune status were as follows: HAI versus FIA, 95.3% (283 of 297); HAI versus PLA (1:10 dilution), 96.3% (286 of 297); HAI versus PLA (undiluted), 93.9% (279 of 297); PLA (1:10 dilution) versus FIA, 94.9% (282 of 297); and PLA (undiluted) versus FIA, 97.9% (291 of 297). The HAI test is the most time consuming, subjective, and technically difficult to perform. The FIA and PLA tests are very rapid and less labor intensive. In addition, the FIA offers an objective determination of the patient's rubella antibody level.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=272717Documentos Relacionados
- Comparison of enzyme-linked immunosorbent assay, hemagglutination inhibition, and passive latex agglutination for determination of rubella immune status.
- Comparison of Enzyme-Linked Immunosorbent Assay, Hemagglutination Inhibition, and Passive Latex Agglutination for Determination of Rubella Immune Status
- New latex agglutination test for rapid determination of rubella immune status.
- Problems in determining immune status in borderline specimens in an enzyme immunoassay for rubella immunoglobulin G antibody.
- Comparison of several test systems used for determination of rubella immune status.