Laboratory evaluation of amikacin susceptibility testing by the AutoBac I system.
AUTOR(ES)
Boughton, W H
RESUMO
A total of 319 clinical isolates were compared for amikacin susceptibility by AutoBac I and Bauer-Kirby disk diffusion methods. There was an overall agreement of 98.5% between AutoBac I and Bauer-Kirby. An additional 23 strains resistant to amikacin, obtained from Bristol Laboratories, registered a 65% overall agreement. The sensitivity readings differed between the two methods for 13 organisms: 11 were minor discrepancies, intermediate by one of the two methods, and 2 were very major discrepancies, resistant by disk diffusion and susceptible by AutoBac I.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=273037Documentos Relacionados
- Evaluation of cefamandole susceptibility testing of Enterobacteriacea by the autobac 1 system.
- Evaluation of the Autobac 1 Susceptibility Testing System in a Clinical Diagnostic Laboratory
- Laboratory evaluation of the Anderson Technical semiautomated susceptibility testing system.
- Serum bactericidal testing with the Autobac system.
- Methicillin-Resistant Staphylococcus aureus Susceptibility Testing by an Automated System, Autobac I