Serum gentamicin assay: A comparison and assessment of different methods
AUTOR(ES)
Phillips, I.
RESUMO
We have compared, in a clinical laboratory, three methods for estimating the concentration of gentamicin in serum. The adenylase method is most accurate, but requires considerable skilled technical time and expensive apparatus. The urease method requires an accurate pH meter but is otherwise inexpensive, but in our hands, although it produces results most rapidly, it also requires considerable technician time and is least accurate. The agar diffusion method requires no expensive apparatus, least technician time, and produces results of acceptable accuracy. It does, however, take longer than the other two methods to produce results.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=478153Documentos Relacionados
- Comparison of five methods for the assay of serum gentamicin.
- Comparative assessment of in vitro inactivation of gentamicin in the presence of carbenicillin by three different gentamicin assay methods.
- Latex agglutination inhibition card test for gentamicin assay: clinical evaluation and comparison with radioimmunoassay and bioassay.
- A comparison of serum folate estimations using two different methods.